Agilent E5025 Cryogen Monitor
5
400/54 Premium Shielded NMR Magnet System Product Overview
85
Errors and Alarms
If the unit detects an error or a low cryogen level, the
display flashes to indicate that attention is required.
The instrument has a measurement cycle period of 5
seconds, at which time it measures the level of liquid
nitrogen and then rewrites the display. Helium level is read
automatically three times daily, at 01:00 hrs, 09:00 hrs, 17:00
hrs. If the helium part of the display flashes a message “c”,
then the monitor requires calibration. In the unlikely event
that you see this message, contact Agilent for assistance.
Any command change takes effect at the end of the next
5- second cycle. Therefore, it can take up to 3 cycles (15
seconds) to accept a new menu value and remeasure. Then
on the final cycle error, conditions are reassessed.
•
Nitrogen errors or alarms cause the left side of the
display to flash.
•
Helium errors or alarms cause the right side of the
display to flash.
•
When the entire display flashes, there is a problem with
both
cryogens.
When the display is flashing an error, the display is
periodically held steady for a few seconds, which allows you
to more easily read the display and access the menus.
Using the menu system causes the flashing to stop. The
flashing resumes when you exit the menu. If you alter a
menu item value that brings the flashing to an end, there is
a delay of at least 10 seconds before the instrument
remeasures and refreshes the display flash status.
If the Helium display is flashing and a parameter was
altered, you must manually perform a Helium read in the
USER menu to force an update of the Helium environment.
Содержание 400/54 Premium Shielded NMR Magnet System
Страница 1: ...Agilent Technologies Agilent 400 54 Premium Shielded NMR Magnet System System Overview ...
Страница 6: ...4 400 54 Premium Shielded NMR Magnet System Product Overview ...
Страница 32: ...30 400 54 Premium Shielded NMR Magnet System Product Overview 2 Installation and Operation Safety ...
Страница 36: ...34 400 54 Premium Shielded NMR Magnet System Product Overview 3 Maintenance Figure 2 Magnet orientation ...
Страница 56: ...54 400 54 Premium Shielded NMR Magnet System Product Overview 3 Maintenance ...
Страница 66: ...64 400 54 Premium Shielded NMR Magnet System Product Overview 4 Reference ...
Страница 99: ......
Страница 100: ...Agilent Technologies Agilent Technologies Inc Printed in USA April 2012 9100380100 9100380100 ...