Maintenance
3
400/54 Premium Shielded NMR Magnet System Product Overview
47
Figure 14
Side view of the helium siphon inserted into both magnet and
transfer Dewar for a helium transfer
14
When full, take the pressure off the transfer Dewar to
stop the transfer.
15
Refit the helium nonreturn valve to the manifold.
Figure 15
Helium siphon in the transfer Dewar showing it lifted from the
base of the Dewar
16
Inspect the bung and baffle assembly for damage and
ensure that it is warm and dry (no water droplets)
before refitting.