Appendix Service Information
Calibration Reference
192
Calibration Reference
Before you calibrate the power supply, you must unsecure it by entering the
correct security code. See “Calibration Overview”, starting on page 64, for
more detailed procedures to unsecure or secure the power supply.
Agilent Technologies Calibration Services
When your power supply is due for calibration, contact your local Agilent
Technologies Service Center for a low-cost calibration. The Agilent E3646A/
47A/48A and E3649A power supplies are supported on calibration processes
which allow Agilent Technologies to provide this service at competitive prices.
Calibration Interval
Recommended calibration interval for this power supply is 1 year. This will
ensure that your power supply will remain within specification for the next
calibration interval. Agilent Technologies does not recommend extending
calibration intervals beyond 1 year for any application. This criteria for re-
adjustment provides the best long-term stability.
To Unsecure the Power Supply Without the Security Code
To unsecure the power supply without the correct security code (when you
forget the security code), follow the steps below. See “Electrostatic Discharge
(ESD) Precautions” on page 170 before beginning this procedure.
1 Disconnect the power cord and all load connections from the power supply.
2 Remove the instrument cover. Refer to the disassembly drawing on page 174.
3 Connect the power cord and turn on the calibration mode by holding down the
Calibrate
key as you turn on the power supply and
hold down
the key until you
hear a long beep. Be careful not to touch the power line connections.
4 Apply a short between the two exposed metal pads on JP3 (located near U17).
The JP3 is outlined with a circle on the Digital board component locator (
top
& bottom side
) drawing.
5 While maintaining the short, move to the security code and enter any unsecure
code in the calibration mode. The power supply is now unsecured.
6 Remove the short at JP3. (An error occurs if not removed.)
7 Turn off and reassemble the power supply.
Now you can enter a new security code. Be sure you take note of the new
security code.
Содержание 2110-0457
Страница 9: ...8...
Страница 15: ...14 Contents Contents...
Страница 16: ...1 Quick Start...
Страница 26: ...2 General Information...
Страница 38: ...3 Front Panel Operation and Features...
Страница 70: ...4 Remote Interface Reference...
Страница 122: ...5 Error Messages...
Страница 133: ...Chapter 5 Error Messages Calibration Errors 132...
Страница 134: ...6 Application Programs...
Страница 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Страница 146: ...7 Tutorial...
Страница 157: ...Chapter 7 Tutorial Remote Programming 156...
Страница 158: ...8 Specifications...
Страница 165: ...Chapter 8 Specifications Supplemental Characteristics 164 Figure 8 2 Dimensions for Rack mounting...
Страница 166: ...Appendix Service Information...
Страница 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Страница 202: ......
Страница 203: ......
Страница 204: ......
Страница 205: ......
Страница 209: ...208 Index Index...