Manual Change
Source Voltage Accuracy Test
Source Voltage Accuracy Test
The 4339B’s source voltage is measured with a DC voltmeter.
Specification
Test Equipment
Procedure
WARNING
When the High Voltage indicator is lit, the 4339B outputs high voltages of up to 1000
Vdc maximum.
Step 1.
Reset the 4339B using the following procedure:
1. Press [blue][
⋅
] to display the system reset menu.
2. Select Yes using [
←
][
↓
]or [
↑
][
→
] and press [Enter].
Step 2.
Set up the equipment as shown in Figure 1.
Source Voltage (Vs) Accuracy:
±
(0.16 % + 100mV) (Vs
≤
200V)
±
(0.16 % + 500mV) (Vs>200V)
Description
Recommended Model
Multimeter
3458A
V Measurement Adapter
PN 04339-65006
Banana-Banana Jumper
11508A