Manual Change
Source Voltage Accuracy Test
Step 12.
Press [Meas Time] to set the measurement time to SHORT. The current measurement time
setting is indicated by the
Meas Time
annunciator.
Step 13.
Press [Trig Mode] to set the trigger mode to Manual. The current trigger mode setting is
indicated by the
Trigger
annunciator.
Step 14.
Press [V Output] to turn the source voltage ON.(V Output indicator lights.)
Step 15.
Press [Trig] to measure.
Step 16.
Record the 4339B reading on the calculation sheet.
Step 17.
Press [V Output] to turn the source voltage OFF. (V Output indicator turns OFF.)
Step 18.
Calculate the test result according to the calculation sheet, and record the result into the
performance test recored.
Step 19.
Perform the test for all settings listed in Table 4. The source voltage must be turned OFF
after each test.
Resistance Measurement Accuracy Test (Grounded Device).
Step 20.
Set up the equipment as shown in Figure 8.
Table 4
Registance Measurement Accuracy Test Settings
Standard
Resistor
Voltage
Setting
Measurement
Time
16353C
100V
Short
16353D
100V
Long
16353E
100V
Long
16353F
100V
Long
16353G
100V
Long
16353H
100V
Long
16353H
100V
Short
16353H
10V
Long