![Agilent Technologies 16353A Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/agilent-technologies/16353a/16353a_manual-change_2867821017.webp)
Manual Change
Source Voltage Accuracy Test
Figure 8
Resistance Measurement Accuracy Test Setup (Grounded)
Step 21.
Set the RC Box Adapter switch to the
G
position.
Step 22.
Set the source voltage to 100V using the following procedure:
1. Press [Source Voltage] to display the source voltage setting.
2. Press [1] [0] [0] [Enter] to set the source voltage to 100V.
3. Confirm that + 100V is displayed on the right side of the LCD.
Step 23.
Press [V Output] to turn the source voltage ON. (V Output indicator turns ON.)
Step 24.
Press [blue] [4]. Select
OpenMeas
and press [Enter] to perform the OPEN correction.
Step 25.
Press [V Output] to turn the source voltage OFF. (V Output indicator turns OFF.)
Step 26.
Connect the 16353D to the RC Box Adapter as shown in Figure 9.
Figure 9
16353x Connection (Grounded)
Step 27.
Press [V Output] to turn the source voltage ON. (V Output indicator turns ON.)
4339boe1008
4339B
: No Connection
RC Box
Adapter
4339boe1005
RC Box Adapter
Direct connection, no cable
BNC Cable
16353x