Agilent Technologies 16353A Скачать руководство пользователя страница 12

 

Manual Change

Source Voltage Accuracy Test

Figure 5 

16353x Connection

Step 13.

Press [MeasPrmtr]. Select I and press [Enter] to set the measurement parameter to current 
(I).

Step 14.

Press [MeasTime] to set the measurement time to LONG. The current measurement time 
setting is indicated by the 

Meas Time

 annunciator.

Step 15.

Press [Trig Mode] to set the trigger mode to 

Manual

. The current trigger mode setting is 

indicated by the 

Trigger

 annunciator.

Step 16.

Set the source voltage to 1V using the following procedure:

1. Press [Source Voltage] to display the source voltage setup menu.

2. Press [1][Enter] to set the source voltage to 1V.

3. Confirm that +1V is displayed on the right side of the LCD.

Step 17.

Press [V Output] to turn the source voltage ON. (V Output indicator turns ON.)

Step 18.

Press [Trig] to measure.

Step 19.

Record the multimeter reading on the calculation sheet on the Multimeter Reading line.

Step 20.

Record the 4339B reading on the calculation sheet on the 4339B Reading line.

Step 21.

Press [V Output] to turn the source voltage OFF. (V Output indicator turns OFF.)

Step 22.

Calculate the test result according to the calculation sheet, and record the result into the 
performance test record.

Step 23.

Perform this test for all settings listed in Table 3. The source voltage must be turned OFF 
after each test. 

4339boe1004

RC Box Adapter

Direct connection, no cable

BNC Cable

16353x

Содержание 16353A

Страница 1: ...e describes the procedure when use the Agilent 16353A 16353H Standard Resistor Set instead of the Agilent 16340A RC Box in 4339B Performance Test When use the Agilent 16340A RC Box refer to the Chapte...

Страница 2: ...x Adapter V Measurement Adapter BNC F TXA Ground Open F Adapter BNC F F Adapter Cable Cable Handler Interface Tester Voltage Range 0V to 1000V Accuracy 0 04 No substitute No substitute No substitute B...

Страница 3: ...e tests described in this section Use only calibrated test instruments when performance testing the 4339B4 Equipment which equals or surpasses the key required specifications of the recommended equipm...

Страница 4: ...for the recommended test equipment Test results recorded during incoming inspection can be used for comparison purposes during periodic maintenance troubleshooting and after repair or adjustment Calib...

Страница 5: ...e 4339B outputs high voltages of up to 1000 Vdc maximum Step 1 Reset the 4339B using the following procedure 1 Press blue to display the system reset menu 2 Select Yes using or and press Enter Step 2...

Страница 6: ...he source voltage OFF V Ouptut indicator turns OFF Step 8 Calculate the test result according to the calculation sheet and record the result into the performance test record Step 9 Perform this test f...

Страница 7: ...Manual Change Source Voltage Accuracy Test 3 The desired voltage will be displayed Table 2 Source Voltage Accuracy Test Settings Source Voltage Setting 0V 10V 25V 50V 100V 200V 201V 250V 500V 1000V...

Страница 8: ...re 1 Press blue to display the system reset menu 2 Select Yes using or and press Enter Step 2 Set up the equipment as shown in Figure 2 Figure 2 Ammeter Offset Voltage Test Setup Ammeter Offset Voltag...

Страница 9: ...nput resistance test Ammeter Input Resistance test Step 9 Record the 42038A 10 k calibration value and the ammeter offset voltage test result on the calculation sheet Step 10 Connect the 42038A to the...

Страница 10: ...values on the calculation sheet Step 2 Reset the 4339B using the following procedure 1 Press blue to display the system reset menu 2 Select Yes using or and press Enter Step 3 Set the Offset error Ca...

Страница 11: ...Source Voltage to display the source voltage setup menu 2 Press 1 0 Enter to set the source voltage to 10V 3 Confirm that 10V is displayed on the right side of the LCD Step 9 Press V Output to turn t...

Страница 12: ...ter to set the source voltage to 1V 3 Confirm that 1V is displayed on the right side of the LCD Step 17 Press V Output to turn the source voltage ON V Output indicator turns ON Step 18 Press Trig to m...

Страница 13: ...Current Measurement Accuracy Test Settings Test Current Standard Resistor Voltage Setting Measurement Time 100pA 16353G 1V LONG 1nA 16353F 1V LONG 10nA 16353E 1V LONG 100nA 16353D 1V LONG 1 A 16353C...

Страница 14: ...calibration values into the Calculation Sheet Step 2 Reset the 4339B using the following procedure 1 Press blue to display the system reset menu 2 Select Yes using or and press Enter Step 3 Set the O...

Страница 15: ...1 0 0 Enter to set the source voltage to 100V 3 Confirm that 100V is displayed on the right side of the LCD Step 8 Press V Output to turn the source voltage ON V Output indicator turns ON Step 9 Press...

Страница 16: ...sheet Step 17 Press V Output to turn the source voltage OFF V Output indicator turns OFF Step 18 Calculate the test result according to the calculation sheet and record the result into the performanc...

Страница 17: ...the right side of the LCD Step 23 Press V Output to turn the source voltage ON V Output indicator turns ON Step 24 Press blue 4 Select OpenMeas and press Enter to perform the OPEN correction Step 25...

Страница 18: ...Output indicator turns OFF Step 31 Calculate the test result according to the calculation sheet and record the result into the performance test record Step 32 Perform this test for all settings liste...