Manual Change
Source Voltage Accuracy Test
Figure 5
16353x Connection
Step 13.
Press [MeasPrmtr]. Select I and press [Enter] to set the measurement parameter to current
(I).
Step 14.
Press [MeasTime] to set the measurement time to LONG. The current measurement time
setting is indicated by the
Meas Time
annunciator.
Step 15.
Press [Trig Mode] to set the trigger mode to
Manual
. The current trigger mode setting is
indicated by the
Trigger
annunciator.
Step 16.
Set the source voltage to 1V using the following procedure:
1. Press [Source Voltage] to display the source voltage setup menu.
2. Press [1][Enter] to set the source voltage to 1V.
3. Confirm that +1V is displayed on the right side of the LCD.
Step 17.
Press [V Output] to turn the source voltage ON. (V Output indicator turns ON.)
Step 18.
Press [Trig] to measure.
Step 19.
Record the multimeter reading on the calculation sheet on the Multimeter Reading line.
Step 20.
Record the 4339B reading on the calculation sheet on the 4339B Reading line.
Step 21.
Press [V Output] to turn the source voltage OFF. (V Output indicator turns OFF.)
Step 22.
Calculate the test result according to the calculation sheet, and record the result into the
performance test record.
Step 23.
Perform this test for all settings listed in Table 3. The source voltage must be turned OFF
after each test.
4339boe1004
RC Box Adapter
Direct connection, no cable
BNC Cable
16353x