Error Information
General Error Messages
Leak
Error ID: 0064
A leak was detected in the module.
The signals from the two temperature sensors (leak sensor and board-mounted
temperature-compensation sensor) are used by the leak algorithm to determine
whether a leak is present. When a leak occurs, the leak sensor is cooled by the
solvent. This changes the resistance of the leak sensor which is sensed by the
leak-sensor circuit on the mainboard.
Open Cover
Error ID: 0205
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the foam
is removed, the fan is switched off, and the error message is generated.
Probable cause
Suggested actions
1
Loose fittings.
Ensure all fittings are tight.
2
Broken capillary.
Exchange defective capillaries.
3
Leaking valve.
Exchange valve.
4
Leaking flow cell.
Exchange optical unit.
Probable cause
Suggested actions
1
The top foam was removed during operation. Please contact your Agilent service representa-
tive.
2
Foam not activating the sensor.
Please contact your Agilent service representa-
tive.
3
Defective sensor or main board.
Please contact your Agilent service representa-
tive.