Agilent InfinityLab LC Series RID User Manual
39
2
Site Requirements and Specifications
Performance Specifications
Safety and maintenance
Extensive diagnostics, error detection and display, leak detection,
safe leak handling, leak output signal for shutdown of pumping sys-
tem. Low voltages in major maintenance areas.
GLP
Early maintenance feedback (EMF) for continuous tracking of instru-
ment usage with user-selectable limits and feedback messages. Elec-
tronic records of maintenance and errors. Automated operational
qualification/performance verification (OQ/PV).
Housing
All materials recyclable
Table 2
Agilent 1260 Infinity II Refractive Index Detector (G7162A) Performance Specifications
Type
Specification
NOTE
Based on ASTM method E-1303-95 Practice for Refractive Index Detectors used
in Liquid Chromatography. Reference conditions; optics temperature 35 °C,
response time 4 s, flow 1.0 mL/min LC-grade Water, restriction capillary, column
compartment temperature 35 °C, Agilent on-line degasser (e.g. G4225A), pump
and thermostatted column compartment. Instrument equilibrated for 2 hours.