Agilent InfinityLab LC Series RID User Manual
95
7
Error Information
Refractive Index Detector Specific Error Messages
Heater Fuse
Error ID: 2682
The electronic fuse of the heater has been activated.
Wrong Temperature Profile
Error ID: 2683
After turning ON the optical unit heat control, the temperature does not increase
at a sufficiently fast rate to reach the set point.
Probable cause
Suggested actions
1
Short in heater circuit.
Powercycle the module.
2
Defective main board.
Please contact your Agilent service representa-
tive.
3
Defective heater.
Please contact your Agilent service representa-
tive.
Probable cause
Suggested actions
1
Defective main board.
Please contact your Agilent service representa-
tive.
2
Defective heater.
Please contact your Agilent service representa-