70
Chapter 3
Performance Tests
AVERAGE NOISE LEVEL TEST
P
e
rf
orm
a
nc
e T
e
s
ts
CAUTION
Make sure the 8349A Amplifier, used in the signal generator system for
11970A, Q and U tests, is set for external leveling before you turn it on.
Failure to set this amplifier for external leveling may allow the amplifier
output to rise about 20 dBm, which is high enough to damage the frequency
tripler.
6. Disconnect the power sensor and connect the equipment as shown in
Figure 3-1
.
7. On the 8566A/B Spectrum Analyzer, set the controls as follows:
For 11970K, A, Q or U:
SHIFT
,
DATA STEP
[
↑
]
KSU
FREQUENCY SPAN
3,
0
,
MHz
VIDEO BW
,
1,
0
,
0
,
Hz
RES BW,
1
,
0,
0,
kHz
For 11970V or W:
SHIFT
,
DATA STEP
[
↑
]
KSU
FREQUENCY SPAN
,
1,
0,
0
,
MHz
VIDEO BW,
3
,
MHz
RES BW
,
3
,
MHz
and
CENTER FREQUENCY
:
For 11970K:
2,
2,
GHz
For 11970A:
3
,
3
,
GHz
For 11970Q:
4
,
1
,
GHz
For 11970U:
5,
0
,
GHz
For 11970V:
6
,
2
,[
.
]
5
,
GHz
For 11970W:
9,
2
,[
.
]
5
,
GHz
8. Press
SHIFT
and SWEEP,
CONT
,
KSt
.
9. Press
CENTER FREQUENCY
again and enter the appropriate center frequency as follows:
For 11970K:
1
,
8
,
GHz
For 11970A:
2
,
6
, [
.
]
5
,
GHz
For 11970Q:
3,
3,
GHz
For 11970U:
4
,
0
,
GHz
Содержание 11970 Series
Страница 4: ...4 ...
Страница 7: ...7 General Information 1 General Information ...
Страница 16: ...16 Chapter1 General Information Specifications General Information ...
Страница 17: ...17 Operation 2 Operation ...
Страница 25: ...Chapter 2 25 Operation Using the Mixers with the E4407B Spectrum Analyzer Option AYZ Operation Figure 2 4 ...
Страница 49: ...49 Performance Tests 3 Performance Tests ...
Страница 73: ...73 Service 4 Service ...
Страница 78: ...78 Chapter4 Service Maintenance Service ...