Chapter 3
55
Performance Tests
Performance Test Procedures
P
e
rf
orm
a
nce
T
est
s
Table 3-5. Recommended Test Equipment for 11970V
Instrument
Critical Specifications
Recommended Model
Spectrum Analyzer
LO and IF ranges compatible with mixer
8566B
Swept Source
Compatible with V-Band Plug-In
8350B
V-Band Source Plug-In
Frequency Range: 50 to 75 GHz
Output Level: >0 dBm
Hughes
1
47724H-1510
1. Hughes Aircraft Co. Electron Dynamics Division, Torrance, CA
Amplifier
Output Level >18 dBm leveled
Frequency Range: 4 to 6 GHz
11975A
Power Meter
Compatible with Power Sensor
432A
Thermistor Mount
SWR: <2.0
Hughes 45774H-1100
Thermistor Mount
SWR: <1.3 @ 6 GHz
478A
Directional Coupler
2
2. Calibration data for the coupling ration between the output arm and the auxiliary
arm is necessary for accurate measurements.
Coupling: 20 dB
Directivity: >20 dB
Auxiliary Arm SWR: <1.5
Hughes 45324H-1220
Variable Attentuator
Range: 0 to 15 dB
TRG
3
510V/385
3. Alpha Industries Inc. TRG Division, Woburn, MA
Adapter
SMA female to Type N female
P/N 1250-1404
Adapter
SMA female to SMA female
P/N 1250-1158
Adapter (2 required)
Type N male to SMA female
P/N 1250-1250
Cables (4 required)
Connectors: SMA male
P/N 5061-5458
Cable
Connectors: SMA male
Loss: <1.0 dB @ 20 GHz
P/N 8120-4396
Cable
Connectors: BNC male
11086A
Содержание 11970 Series
Страница 4: ...4 ...
Страница 7: ...7 General Information 1 General Information ...
Страница 16: ...16 Chapter1 General Information Specifications General Information ...
Страница 17: ...17 Operation 2 Operation ...
Страница 25: ...Chapter 2 25 Operation Using the Mixers with the E4407B Spectrum Analyzer Option AYZ Operation Figure 2 4 ...
Страница 49: ...49 Performance Tests 3 Performance Tests ...
Страница 73: ...73 Service 4 Service ...
Страница 78: ...78 Chapter4 Service Maintenance Service ...