Performance Tests
Conversion Loss and Frequency Response
P
e
rf
orm
a
nc
e T
e
s
ts
9. Adjust the output power of the signal generator for a reading of approximately -10 dBm
on the power meter for the 11970K, V or W and for approximately -3 dBm for the
11970A, Q or U. (Make sure that the unleveled light is not on for the 11970V and
11970W. Vary the power by adjusting the rotary vane attenuator.)
10.On the spectrum analyzer, press
SHIFT
and [
MKR/
∆
⇑
STP SIZE
].
11.Set spectrum analyzer
CENTER FREQUENCY
to:
11970K: 22 GHz
11970A: 33 GHz
11970Q: 41 GHz
11970U: 50 GHz
11970V: 62 GHz
11970W: 92 GHz
12.On the spectrum analyzer, press
SHIFT
and
SWEEP,
CONT
.
13.Set the spectrum analyzer controls as follows:
For 11970K, A, Q or U:
SHIFT
,
REFERENCE LEVEL
,
3
,
0
,
-dB
SCALE
ENTER dB/DIV
,
2
,
dB
RES BW
,
1
,
MHz
FREQUENCY SPAN
,
2
,
0
,
MHz
CF STEP SIZE
,
5
,
0
,
0
,
MHz
For 11970V or W:
SHIFT
,
REFERENCE LEVEL,
3
,
0
,
-dB
SCALE
ENTER dB/DIV
,
2
,
dB
RES BW
,
3,
MHz
FREQUENCY SPAN
,
2
,
0
,
0
,
MHz
CF STEP SIZE
,
1
,
GHz
Press
REFERENCE LEVEL
:
11970K:
3
,
6
,
-dBm
11970A:
3
,
8
,
-dBm
11970Q:
4
,
0
,
-dBm
11970U:
4
,
0
,
-dBm
11970V:
4
,
8
,
-dBm
11970W:
5
,
6
,
-dBm
Содержание 11970 Series
Страница 4: ...4 ...
Страница 7: ...7 General Information 1 General Information ...
Страница 16: ...16 Chapter1 General Information Specifications General Information ...
Страница 17: ...17 Operation 2 Operation ...
Страница 25: ...Chapter 2 25 Operation Using the Mixers with the E4407B Spectrum Analyzer Option AYZ Operation Figure 2 4 ...
Страница 49: ...49 Performance Tests 3 Performance Tests ...
Страница 73: ...73 Service 4 Service ...
Страница 78: ...78 Chapter4 Service Maintenance Service ...