Chapter 2
Operation
Using the Mixers with the 856X Series Spectrum Analyzers
O
p
er
ati
o
n
Identify signals in wide frequency spans
7.
SIG ID AT MKR
identifies signals in wide frequency spans, using harmonic search.
SIG ID
AT MKR
automatically determines the proper frequency of a signal and displays its
value on the spectrum analyzer.
• Activating
SIG ID AT MKR
on an image of the signal will yield a reading in the active
block, as shown in
Figure 2-14
. The true signal frequency is given with an identifier
that the marker resides on an image. If the marker is placed on a true signal, and
SIG ID AT MKR
is activated, the signal frequency will appear without the
IMAGE
notation, as shown in
Figure 2-15
.
8. To activate
SIG ID AT MKR
:
a. Place a marker on a signal.
b. Press
AUX CTRL
,
EXTERNAL MIXER
,
SIGNAL IDENT
,
SIG ID AT MKR
.
9. To exit the external mixer mode, press
AUX CTRL
, then
INTERNAL MIXER
.
Figure 2-14
SIG ID AT MKR
Performed on an Image Signal
Содержание 11970 Series
Страница 4: ...4 ...
Страница 7: ...7 General Information 1 General Information ...
Страница 16: ...16 Chapter1 General Information Specifications General Information ...
Страница 17: ...17 Operation 2 Operation ...
Страница 25: ...Chapter 2 25 Operation Using the Mixers with the E4407B Spectrum Analyzer Option AYZ Operation Figure 2 4 ...
Страница 49: ...49 Performance Tests 3 Performance Tests ...
Страница 73: ...73 Service 4 Service ...
Страница 78: ...78 Chapter4 Service Maintenance Service ...