Remote Interface Configurations
7
11713B/C Operating and Service Manual
83
3
Manually
enter IP
ADDRESS,
SUBNET and
GATEWAY
a
At IO SETUP screen,
press
IP ADDR
softkey
b
At IP ADDRESS
screen, press
MANUAL IP
softkey
c
At MANUAL IP
ADDRESS screen,
enter
IP ADDRESS
,
SUBNET
and
GATEWAY
(scroll
using
NEXT ITEM
softkey)
d
Then
SAVE
.
e
Press
BACK
softkey to
go back to IP
ADDRESS screen
4
Verify LAN
status on
11713B/C.
a
At IO SETUP screen,
select
IP ADDR
softkey
b
At IP ADDRESS
screen, select
LAN
STAT
softkey
c
Verify values
CURRENT IP
ADDRESS, CURRENT
SUBNET and
CURRENT GATEWAY
are valid
d
Ensure LAN STATUS
is
NORMAL
e
If LAN STATUS is
FAULT,
check your
LAN connection and
then press
Restart
Netw
softkey and wait
f
Press
BACK
softkey to
go back to IP
ADDRESS screen
Step
Action
Illustration
Содержание 11713B
Страница 1: ...Agilent Technologies Agilent 11713B C Attenuator Switch Drivers Operating and Service Manual ...
Страница 12: ...12 11713B C Operating and Service Manual ...
Страница 24: ...24 11713B C Operating and Service Manual 1 Introduction ...
Страница 43: ...Specifications 3 11713B C Operating and Service Manual 43 Declaration of Conformity ...
Страница 44: ...44 11713B C Operating and Service Manual 3 Specifications ...
Страница 64: ...64 11713B C Operating and Service Manual 5 Local Operations ...
Страница 88: ...88 11713B C Operating and Service Manual 7 Remote Interface Configurations ...
Страница 104: ...104 11713B C Operating and Service Manual 8 Replacing Major Assemblies ...
Страница 114: ...114 11713B C Operating and Service Manual 9 Servicing the Attenuator Switch Driver ...