48
11713B/C Operating and Service Manual
4
Verification
Operator’s Check for Remote Operation
These procedure verify that the driver can be controlled remotely
using GPIB, USB and/or LAN.
1
Refer to
, “Remote Interface Configurations, to connect
11713B/C to your computer through GPIB, USB and/or LAN.
2
Once remote connection is available, send the following SCPI
commands to the driver and note the changes on front panel
LEDs.
If the above checks are successful, the driver’s remote operation is
working correctly. These procedures do not check all of the drivers’s
program codes that can be executed. However, if the driver work
correctly from the front panel, there is a high probability that the
driver will respond to all the program codes.
Description
Command
Observation
Close switching paths from
channel 1 to channel 4 (bank1)
ROUTe:CLOSe (@101:104)
LEDs for pushbuttons 1 to
4 light ON
Open switching paths from
channel 5 to channel 8 (bank 1)
ROUTe:OPEn (@105:108)
LEDs for pushbuttons 5 to
8 light OFF
Query status on channel 2 (bank
1)
ROUTe:CLOSe? (@102)
Return value “1” (LED for
pushbutton 2 light ON)
Query status on channel 7 (bank
1)
ROUTe:CLOSe? (@107)
Return value “0” (LED for
pushbutton 7 light OFF)
Содержание 11713B
Страница 1: ...Agilent Technologies Agilent 11713B C Attenuator Switch Drivers Operating and Service Manual ...
Страница 12: ...12 11713B C Operating and Service Manual ...
Страница 24: ...24 11713B C Operating and Service Manual 1 Introduction ...
Страница 43: ...Specifications 3 11713B C Operating and Service Manual 43 Declaration of Conformity ...
Страница 44: ...44 11713B C Operating and Service Manual 3 Specifications ...
Страница 64: ...64 11713B C Operating and Service Manual 5 Local Operations ...
Страница 88: ...88 11713B C Operating and Service Manual 7 Remote Interface Configurations ...
Страница 104: ...104 11713B C Operating and Service Manual 8 Replacing Major Assemblies ...
Страница 114: ...114 11713B C Operating and Service Manual 9 Servicing the Attenuator Switch Driver ...