Remote Operations
6
11713B/C Operating and Service Manual
71
Commands Relevant to Switches
To Open/Close Switch
The commands below close/open switching path(s).
•
[:ROUTe]:CLOSe <channel list>
•
[:ROUTe]:OPEn <channel list>
•
[:ROUTe]:CLOSe:ALL
•
[:ROUTe]:OPEn:ALL
To Query Switch Status
The commands below query the status of switching path(s). Value to
be returned: “1” if true and “0” if false.
•
[:ROUTe]:CLOSe? <channel list>
•
[:ROUTe]:OPEn? <channel list>
To Set Supply Voltage
The commands below set the supply voltage for each bank.
•
CONFigure:BANK1 {OFF|P5v|P15v|P24v|USER}
•
CONFigure:BANK2 {OFF|P5v|P15v|P24v|USER}
To Query Supply Voltage
The commands below query the supply voltage set for each bank.
Value to be returned: OFF, P5, P15, P24 or USER.
•
CONFigure:BANK1?
•
CONFigure:BANK2?
To Set TTL ON/OFF
The commands below set TTL ON/OFF for each bank.
•
CONFigure:BANK1:TTL {OFF|ON}
•
CONFigure:BANK2:TTL {OFF|ON}
To Query TTL status
The commands below query the supply voltage status for each bank.
Value to be returned: “1” if true and “0” if false.
•
CONFigure:BANK1:TTL?
•
CONFigure:BANK2:TTL?
Содержание 11713B
Страница 1: ...Agilent Technologies Agilent 11713B C Attenuator Switch Drivers Operating and Service Manual ...
Страница 12: ...12 11713B C Operating and Service Manual ...
Страница 24: ...24 11713B C Operating and Service Manual 1 Introduction ...
Страница 43: ...Specifications 3 11713B C Operating and Service Manual 43 Declaration of Conformity ...
Страница 44: ...44 11713B C Operating and Service Manual 3 Specifications ...
Страница 64: ...64 11713B C Operating and Service Manual 5 Local Operations ...
Страница 88: ...88 11713B C Operating and Service Manual 7 Remote Interface Configurations ...
Страница 104: ...104 11713B C Operating and Service Manual 8 Replacing Major Assemblies ...
Страница 114: ...114 11713B C Operating and Service Manual 9 Servicing the Attenuator Switch Driver ...