11645A
4-3
Performance Verification
Recertification
Recertification
The following will be provided with a recertified kit:
• a new calibration sticker affixed to the case
• a certificate of calibration
• a calibration report for each device in the kit listing measured values, specifications,
and uncertainties
NOTE
A list of NIST traceable numbers may be purchased upon request to be
included in the calibration report.
Agilent Technologies offers a Standard calibration for the recertification of the kit. For
more information, contact Agilent Technologies. Refer to
“Contacting Agilent” on page 5-5
.
How Often to Recertify
The suggested initial interval for recertification is 12 months or sooner. The actual need for
recertification depends on the use of the kit. After reviewing the results of the initial
recertification, you may establish a different recertification interval that reflects the usage
and wear of the kit.
NOTE
The recertification interval should begin on the date the kit is first used after
the recertification date.
Where to Send a Kit for Recertification
Contact Agilent Technologies for information on where to send your kit for recertification.
See
“Contacting Agilent” on page 5-5
. Refer to
“Returning a Kit or Device to Agilent
for details on sending your kit.
Содержание 11645A Series
Страница 4: ...iv ...
Страница 6: ...vi 11645A Contents ...
Страница 7: ...11645A 1 1 1 General Information ...
Страница 11: ...11645A 2 1 2 Specifications ...
Страница 14: ...2 4 11645A Specifications Electrical Specifications ...
Страница 15: ...11645A 3 1 3 Use Maintenance and Care of the Devices ...
Страница 22: ...3 8 11645A Use Maintenance and Care of the Devices Handling and Storage ...
Страница 23: ...11645A 4 1 4 Performance Verification ...
Страница 26: ...4 4 11645A Performance Verification Recertification ...
Страница 27: ...11645A 5 1 5 Troubleshooting ...
Страница 29: ...11645A 5 3 Troubleshooting Troubleshooting Process Figure 5 1 Troubleshooting Flowchart ...
Страница 32: ...5 6 11645A Troubleshooting Contacting Agilent ...
Страница 33: ...11645A 6 1 6 Replaceable Parts ...
Страница 35: ...11645A 6 3 Replaceable Parts Introduction Figure 6 1 R Band Component Identification Sheet ...
Страница 37: ...11645A 6 5 Replaceable Parts Introduction Figure 6 2 Q Band Component Identification Sheet ...
Страница 39: ...11645A 6 7 Replaceable Parts Introduction Figure 6 3 U Band Component Identification Sheet ...
Страница 41: ...11645A 6 9 Replaceable Parts Introduction Figure 6 4 V Band Component Identification Sheet ...
Страница 43: ...11645A 6 11 Replaceable Parts Introduction Figure 6 5 W Band Component Identification Sheet ...