2-2
11645A
Specifications
Environmental Requirements
Environmental Requirements
Temperature—What To Watch Out For
Due to the small dimensions of the devices, electrical characteristics will change with
temperature. Therefore, the operating temperature is a critical factor in their performance,
and must be stable before use.
IMPORTANT
Avoid unnecessary handling of the devices during use because your fingers
act as a heat source and may increase the temperature of the device.
Table 2-1
Environmental Requirements
Parameter
Required Values/Ranges
Temperature
Operating
a
a. The temperature range over which the calibration standards maintain conformance to their
specifications.
+20
°
C to +26
°
C (+68
°
F to +79
°
F)
Error-corrected range
b
b. The allowable network analyzer ambient temperature drift during measurement calibration and
during measurements when the network analyzer error correction is turned on. Also, the range
over which the network analyzer maintains its specified performance while correction is turned on.
±
1
°
C (1.8
°
F) of measurement calibration temperature
Altitude
Operating
< 4,500 meters (
≈
15,000 feet)
Storage
< 15,000 meters (
≈
50,000 feet)
Relative humidity
Always non–condensing
Operating
0 to 80% (26
°
C maximum dry bulb)
Storage
0 to 90%
Содержание 11645A Series
Страница 4: ...iv ...
Страница 6: ...vi 11645A Contents ...
Страница 7: ...11645A 1 1 1 General Information ...
Страница 11: ...11645A 2 1 2 Specifications ...
Страница 14: ...2 4 11645A Specifications Electrical Specifications ...
Страница 15: ...11645A 3 1 3 Use Maintenance and Care of the Devices ...
Страница 22: ...3 8 11645A Use Maintenance and Care of the Devices Handling and Storage ...
Страница 23: ...11645A 4 1 4 Performance Verification ...
Страница 26: ...4 4 11645A Performance Verification Recertification ...
Страница 27: ...11645A 5 1 5 Troubleshooting ...
Страница 29: ...11645A 5 3 Troubleshooting Troubleshooting Process Figure 5 1 Troubleshooting Flowchart ...
Страница 32: ...5 6 11645A Troubleshooting Contacting Agilent ...
Страница 33: ...11645A 6 1 6 Replaceable Parts ...
Страница 35: ...11645A 6 3 Replaceable Parts Introduction Figure 6 1 R Band Component Identification Sheet ...
Страница 37: ...11645A 6 5 Replaceable Parts Introduction Figure 6 2 Q Band Component Identification Sheet ...
Страница 39: ...11645A 6 7 Replaceable Parts Introduction Figure 6 3 U Band Component Identification Sheet ...
Страница 41: ...11645A 6 9 Replaceable Parts Introduction Figure 6 4 V Band Component Identification Sheet ...
Страница 43: ...11645A 6 11 Replaceable Parts Introduction Figure 6 5 W Band Component Identification Sheet ...