Choosing the Sampling Mode
The osilloscope can operate in real- time or equivalent- time sampling
modes.
You can choose the oscilloscope’s sampling mode in the Acquire menu
(accessed by pressing the
[Acquire]
front panel button).
To select the real-time sampling mode
In the real- time sampling mode, single waveforms are sampled at
uniformly spaced intervals. See
Use the real- time sampling mode when capturing (non- repetitive)
single- shot or pulse waveforms.
The 1000B Series oscilloscopes provide real- time sampling rates up to
500 MSa/s (when two channels are on) or 1 GSa/s (when one channel is
on).
Figure 28
Real-Time Sampling Mode
1
1
1
1
N = Waveform Sample is From
Trigger Time
Содержание 1000 Series
Страница 1: ...s1 Agilent 1000B Series Oscilloscopes User s Guide ...
Страница 12: ...12 Agilent 1000B Series Oscilloscopes User s Guide Contents Safety Symbols 150 Index 151 ...
Страница 16: ...16 Agilent 1000B Series Oscilloscopes User s Guide Tables ...
Страница 36: ...1 Getting Started 36 Agilent 1000B Series Oscilloscopes User s Guide ...
Страница 53: ...Displaying Data 2 Agilent 1000B Series Oscilloscopes User s Guide 53 Figure 21 Waveform After Inversion ...
Страница 64: ...2 Displaying Data 64 Agilent 1000B Series Oscilloscopes User s Guide ...
Страница 98: ...3 Capturing Data 98 Agilent 1000B Series Oscilloscopes User s Guide ...
Страница 112: ...4 Making Measurements 112 Agilent 1000B Series Oscilloscopes User s Guide Figure 46 Cursors Key ...
Страница 116: ...4 Making Measurements 116 Agilent 1000B Series Oscilloscopes User s Guide ...
Страница 132: ...6 Oscilloscope Utility Settings 132 Agilent 1000B Series Oscilloscopes User s Guide Figure 53 Utility Key ...