X-Y Format
This format compares the voltage level of two waveforms point by point. It
is useful for studying phase relationships between two waveforms. This
format only applies to channels 1 and 2. Choosing the X- Y display format
displays channel 1 on the horizontal axis and channel 2 on the vertical
axis.
The oscilloscope uses the untriggered sample acquisition mode and
waveform data is displayed as dots. The sampling rate can vary from
4 kSa/s to 100 MSa/s, and the default sampling rate is 1 MSa/s.
Roll
In Roll mode, the waveform display rolls from right to
left, and the minimum horizontal scale setting is
500 ms/div. No trigger or horizontal position control is
available. Roll mode is used in applications similar to the
ones for which Slow Scan mode is used (see
Figure 14
X-Y Display Format Showing Out-of-Phase Waveforms
Содержание 1000 Series
Страница 1: ...s1 Agilent 1000B Series Oscilloscopes User s Guide ...
Страница 12: ...12 Agilent 1000B Series Oscilloscopes User s Guide Contents Safety Symbols 150 Index 151 ...
Страница 16: ...16 Agilent 1000B Series Oscilloscopes User s Guide Tables ...
Страница 36: ...1 Getting Started 36 Agilent 1000B Series Oscilloscopes User s Guide ...
Страница 53: ...Displaying Data 2 Agilent 1000B Series Oscilloscopes User s Guide 53 Figure 21 Waveform After Inversion ...
Страница 64: ...2 Displaying Data 64 Agilent 1000B Series Oscilloscopes User s Guide ...
Страница 98: ...3 Capturing Data 98 Agilent 1000B Series Oscilloscopes User s Guide ...
Страница 112: ...4 Making Measurements 112 Agilent 1000B Series Oscilloscopes User s Guide Figure 46 Cursors Key ...
Страница 116: ...4 Making Measurements 116 Agilent 1000B Series Oscilloscopes User s Guide ...
Страница 132: ...6 Oscilloscope Utility Settings 132 Agilent 1000B Series Oscilloscopes User s Guide Figure 53 Utility Key ...