d-lab.2, d-lab.2plus, netlab.2plus: Preinstallation
Power Supply
AgfaPhoto
2005-05-01 / PN 04100_01
15
5.2
Connection Values, Connection Voltages, Connection Types
We recommend a 3-phase operation, single-phase operation is possible.
CAUTION!
Make sure that the line section of the on-site installation and the connecting
cable is sufficient!
In case of an insufficient cable cross-section the cable get hot: risk of fire.
Connection
type
Pictograph Nominal
supply
voltage
Nominal
frequency
Nominal current
Site connection,
fuse protection
3W + N + PE
220/380 V
230/400 V
240/415 V
50/60 Hz
16 A
3 x 16 A
3W + PE
200 V
210 V
220 V
230 V
240 V
50/60 Hz
25 A
3 x 25 A
2W + PE
1W + N + PE
200 V
210 V
220 V
230 V
240 V
50/60 Hz
32 A
1 x 32 A
CAUTION!
Only the values listed in the table above may be applied between phase and
neutral, or if there is no neutral, between phase and phase.
CAUTION!
For operation with 50 Hz mains: single-phase connection is only permitted for
industrial networks. If the machine is connected to a public low-voltage network
the connection must always be a 3-phase connection (as required by IEC 61000-
3-3).