![AGFA 5420/100 Скачать руководство пользователя страница 48](http://html.mh-extra.com/html/agfa/5420-100/5420-100_user-manual_2866606048.webp)
DX-D 400
User Manual
38
0232F EN
20191213
2.9
QUANTITATIVE INFORMATION
The following tables show the Quantitative Information associated
to this equipment according to the StandardIEC 60601--1--3:2008
and IEC 60601--1--3:2008+AMD1:2013. These tables illustrate
loading factors for image performance and supply Dose indication
examples. Therefore, they are an example of the adjustment of
Loading Factors, Focal Spot Selection, SID and Collimator
opening, which affect to the radiation quality or to the radiation
dose rate applied in normal use.
2.9.1
FUNCTIONAL TESTS PERFORMED TO OBTAIN THE QUANTITATIVE INFORMATION
Equipment:
Rad Positioner with Ralco Collimator.
Instrumentation used:
Dosimeter: Vacudap
Dosimeter: Unfors
Rectangular Phantom made of Polymethyl-methacrylate (PMMA)
layers: 25 cm x 25 cm x 20 cm.
Test Details:
Minimum SID distance: 100 cm.
Maximum SID distance: 180 cm.
Open Collimator size: 13 cm x 13 cm (min.), 43 cm x 43 cm (max.)
The measurements were made with the exposure parameters shown on
the results table:
KVp Range: 40 KVp, 60 KVp, 80 KVp, 100 KVp, 125 KVp
mAs Range: 1 mAs, 2 mAs, 10 mAs, 50 mAs, 100 mAs
Performed measurements of Air Kerma or Air Kerma Rate at the
following designated positions:
-- Distance SID doses
-- Patient (Phantom) Entrance doses and Entrance doses Rate
-- Patient (Phantom) Output doses and Output doses Rate
-- Collimator Output doses
Note
.
Содержание 5420/100
Страница 1: ...DX D 400 User Manual type 5420 100 Technical Publication 0232F EN 20191213 ...
Страница 2: ......
Страница 10: ...DX D 400 User Manual vi 0232F EN 20191213 ...
Страница 18: ...DX D 400 User Manual 8 0232F EN 20191213 This page intentionally left blank ...
Страница 56: ...DX D 400 User Manual 46 0232F EN 20191213 This page intentionally left blank ...
Страница 60: ...DX D 400 User Manual 50 0232F EN 20191213 This page intentionally left blank ...
Страница 104: ...DX D 400 User Manual 94 0232F EN 20191213 This page intentionally left blank ...
Страница 132: ...DX D 400 User Manual 122 0232F EN 20191213 This page intentionally left blank ...
Страница 154: ...DX D 400 User Manual 144 0232F EN 20191213 This page intentionally left blank ...
Страница 180: ...DX D 400 User Manual Appendix B B 8 APR MATRIX RUSSIAN VERSION ...
Страница 187: ......