![AGFA 5420/100 Скачать руководство пользователя страница 157](http://html.mh-extra.com/html/agfa/5420-100/5420-100_user-manual_2866606157.webp)
DX-D 400
User Manual
0232F EN
20191213
147
SECTION 9
TECHNICAL SPECIFICATIONS
9.1
ENVIRONMENTAL REQUIREMENTS
ATMOSPHERIC PRESSURE (hPa)
RELATIVE HUMIDITY (%)
AMBIENT TEMPERATURE
MIN
MAX
MIN
MAX
MIN
MAX
WORKING
700 hPa
1060 hPa
30%
75%
10 oC (50 oF)
35 oC (95 oF)
TRANSPORT & STORAGE
500 hPa
1060 hPa
10%
90%
-20 oC (--4 oF)
70 oC (158 oF)
These environmental conditions do not include the Digital
Detector. Refer to the Digital Detector Documentation.
9.2
X-RAY SYSTEM POSITIONERS
9.2.1
POWER LINE REQUIREMENTS
EQUIPMENT
VOLTAGE
FREQUENCY
MINIMUM INPUT
POWER
REQUIRED
POWER
OUTPUT
BRAKE POWER
SUPPLY
RAD TABLE -- ELEVATING TABLE
110 - 230 VAC,
Single Phase
50/60 Hz
0.6 kVA
0.18 kW
--
RAD TABLE -- FIXED HEIGHT
TABLE
110 - 230 VAC,
Single Phase
50/60 Hz
0.4 kVA
0.18 kW
--
RAD WALL STAND
* See Note
0.5 kVA
0.10 kW (4A)
24 VDC, 2A
*
NOTE: Power Line for the RAD Wall Stand is supplied by the X-ray Generator (110 -- 230 VAC, Single Phase, 50 / 60 Hz)
Note
.
Содержание 5420/100
Страница 1: ...DX D 400 User Manual type 5420 100 Technical Publication 0232F EN 20191213 ...
Страница 2: ......
Страница 10: ...DX D 400 User Manual vi 0232F EN 20191213 ...
Страница 18: ...DX D 400 User Manual 8 0232F EN 20191213 This page intentionally left blank ...
Страница 56: ...DX D 400 User Manual 46 0232F EN 20191213 This page intentionally left blank ...
Страница 60: ...DX D 400 User Manual 50 0232F EN 20191213 This page intentionally left blank ...
Страница 104: ...DX D 400 User Manual 94 0232F EN 20191213 This page intentionally left blank ...
Страница 132: ...DX D 400 User Manual 122 0232F EN 20191213 This page intentionally left blank ...
Страница 154: ...DX D 400 User Manual 144 0232F EN 20191213 This page intentionally left blank ...
Страница 180: ...DX D 400 User Manual Appendix B B 8 APR MATRIX RUSSIAN VERSION ...
Страница 187: ......