Table 5-1
Error Codes
ERROR
DESCRIPTION
WHAT TO DO
-- -- -- -- -- --
on Display
System failure. This indication may appear together with an
error on the Console, and indicates that the error is not
correctable unless the equipment is turned OFF.
Turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E01, E02
Communication error.
Turn the Generator OFF, check the proper external cable connections
and then turn the Generator ON. If the equipment remains inoperative,
turn it OFF and call Field Service.
E03
System failure.
T
th G
t OFF
d ON
E04
The X-ray Generator has activated “
Preparation
” without a
Console command intervention.
Turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E05
External exposure activated during power-up.
Release any external exposure device or buttons.
Turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E06
“
Exposure
” and/or “
Preparation
” orders are activated during
power-up.
Release all the controls.
Turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E07, E08
X-ray Tube configuration error.
Touch the “
Reset
” button.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E09
Generator Overload error. The exposure has been
interrupted because during exposure has been produced
arcing or bad function on the HV circuitry (X-ray Tube, HV
Transformer and/or HV Cables) or a failure of IGBT module
(overheated or defective IGBTs) has been detected.
It can be also shown making a high power and long
exposure with the X-ray tube cool (X-ray Tube has not been
warmed-up).
It is not necessary to touch the “
Reset
” button, the error indication will
disappear automatically.
If the error code persists, turn the Generator OFF and wait 30 minutes
before turning it ON again.
If the equipment remains inoperative, turn it OFF and call Field Service.
E10, E11
System failure.
Touch the “
Reset
” button.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E12
No mA during exposure or mA value is out of range.
Touch the “
Reset
” button.
Repeat with same technique values, If the error code persists try with
E13
No kV during exposure or kV value is out of range.
Repeat with same technique values, If the error code persists try with
another combinations of kV and mA values.
If the equipment remains inoperative, turn it OFF and call Field Service.
E14, E15
System failure.
Touch the “
Reset
” button.
If the error code persists, turn the Generator OFF and ON again.
If the equipment remains inoperative, turn it OFF and call Field Service.
E16
Invalid value of: kV, mA or kW.
Decrease kV, mA or both. touch the “
Reset
” button.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E17
Communication error or system failure.
Turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
E18
Rotor error. The X-ray tube anode is not rotating while
“
Prep
” is active, then exposures are inhibited, or the X-ray
tube anode is rotating without console command.
It is not necessary to touch the “
Reset
” button, the error indication will
disappear automatically.
If the error code persists, turn the Generator OFF and ON.
If the equipment remains inoperative, turn it OFF and call Field Service.
Содержание 5420/100
Страница 1: ...DX D 400 User Manual type 5420 100 Technical Publication 0232F EN 20191213 ...
Страница 2: ......
Страница 10: ...DX D 400 User Manual vi 0232F EN 20191213 ...
Страница 18: ...DX D 400 User Manual 8 0232F EN 20191213 This page intentionally left blank ...
Страница 56: ...DX D 400 User Manual 46 0232F EN 20191213 This page intentionally left blank ...
Страница 60: ...DX D 400 User Manual 50 0232F EN 20191213 This page intentionally left blank ...
Страница 104: ...DX D 400 User Manual 94 0232F EN 20191213 This page intentionally left blank ...
Страница 132: ...DX D 400 User Manual 122 0232F EN 20191213 This page intentionally left blank ...
Страница 154: ...DX D 400 User Manual 144 0232F EN 20191213 This page intentionally left blank ...
Страница 180: ...DX D 400 User Manual Appendix B B 8 APR MATRIX RUSSIAN VERSION ...
Страница 187: ......