Electromagnetic immunity
This DR Detector is intended for operation in the electromagnetic
environment given below. The user of the DR Detector should ensure that it is
used in such an environment.
Resistance to
Jamming Test
IEC 60601 Test
Level
Level of
Agreement
Electromagnetic
Environment
Guidelines
Discharge of
static electricity
in accordance
with
IEC 61000-4-2
± 6 kV contact
discharge
± 8 kV air
discharge
± 6 contact
discharge
± 8 kV air
discharge
Floors should consist
of wood, concrete or
ceramic tiles. The
relative humidity
must be at least 30%,
if the floor is made of
synthetic material.
Fast transient
electrical
disturbance
variables /
bursts in
accordance with
IEC 61000-4-4
± 2 kV for
network leads
± 1 kV for entry
and outlet leads
± 2 kV for
network leads
± 1 kV for entry
and outlet leads
The quality of the
voltage supplied
should correspond to
a typical commercial
or clinical
environment.
Impulse voltages
(surges) in
accordance with
IEC 61000-4-5
± 1 kV push-
pull voltage
± 2 kV common
mode voltage
± 1 kV push-
pull voltage
± 2 kV common
mode voltage
The quality of the
voltage supplied
should correspond to
that of a typical
commercial or
clinical environment.
Voltage
breakthroughs,
short term
interruptions
and variations in
the voltage
supplied in
accordance with
IEC 61000-4-11
< 5% Ur
(> 95%
breakthrough of
Ur) for ½
period
40% Ur (> 60%
breakthrough of
Ur) for 5
periods
70% Ur (30%
breakthrough of
Ur) for 25
periods
< 5% Ur
(> 95%
breakthrough of
Ur) for ½
period
40% Ur (> 60%
breakthrough of
Ur) for 5
periods
70% Ur (30%
breakthrough of
Ur) for 25
periods
The quality of the
voltage supply
should correspond to
that of a typical
commercial or
clinical environment.
If the user wants the
DR Detector to work
continuously, even
when the energy
supply is
interrupted, it is
recommended to use
an energy supply
DR 18M, DR 24M | Remarks for HF-emission and immunity |
75