Starting the DR detector
Note
:
Before operating the detector, start up the NX workstation.
To start the DR Detector:
1.
Check if the DR detector cable is connected to the control unit.
2.
Check if the power cable of the DR detector control unit is connected to
the mains power.
3.
Power on the DR detector using the power switch at the front of the
control unit.
4.
Check the DR Detector status icon on the DR Detector Switch.
The connection indicator is lit. The DR Detector is ready.
Before exposure make sure to check the equipment daily and confirm that it
works properly. Allow the DR detector to warm up for about 10 minutes before
the system is used for clinical purposes. The DR Detector status icon on the DR
Detector switch will show when the DR Detector is ready.