DR 18M, DR 24M
Electrical connection DR Detector
Rated power supply
(power over ethernet)
12V DC
Power consumption
30 W
Environmental conditions (during normal operation)
Room temperature
b5 °C and +35 °C
Maximum temperature change rate
max. 10 °C/hour
Humidity (non condensing)
between 20% and 75% RH
(non-condensing)
Atmospheric pressure
between 500 hPa and 1060 hPa
Environmental conditions (during storage)
Temperature (ambient)
between -10 °C and +40 °C
Maximum temperature change rate
max. 15 °C/hour
Humidity (non condensing)
between 5% and 95%
(non-condensing)
Atmospheric pressure
between 500 hPa and 1060 hPa
Warming-up time
5 minutes minimum
10 minutes recommended
30 minutes before calibration
Dimensions
Dimensions
width x length x height
DR 18M: 267.5 x 194.5 x 14.2 mm
DR 24M: 327.5 x 253.7 x 14.2 mm
Weight
DR 18M: 0.92 kg
DR 24M: 1.20 kg
Chest wall distance
less than 2.0 mm
DR 18M, DR 24M | Technical Data |
67