No image available after exposure
Details
An exposure is performed and no image becomes available on the
NX workstation.
Cause
1.
Due to the automatic exposure control, the dose may become
too low to trigger the DR detector.
2.
The sensor(s) to trigger the automatic exposure detection of
the DR detector are blocked.
Solution Check both of following conditions:
1.
Check the output of the automatic exposure on the X-ray
console. If the mAs value is low to your experience, make sure
to manually raise the mAs (keep other settings) instead of
repeating in automatic mode.
2.
Check if no objects with a high attenuation are blocking part
of the image area.