Environment specifications
Environment requirements
• Ventilated room,
• away from direct sunlight,
• away from sources of dust, humidity, heat and cold,
• room temperature between 15°C (50°F) and 30°C (86°F),
• relative humidity between 20% and 75% noncondensing.
Power requirements
The AC outlet must have either of following specifications:
• 100-120 V, 50-60 Hz, 16/15 A,
• 200-240 V, 50-60 Hz, 16/15 A.
Network requirements
• Ethernet/connectors:
RJ45 twisted pair for 10/100/1000 Base-TX,
• Network protocols (TCP/IP services):
HTTP.
Space requirements
1
2
1.
Space required for servicing and installation
2.
Space required for normal use
156