Remarks for HF-emission and immunity
This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such
an environment.
RF Emission
Measurements
Agree-
ment
Electromagnetic Environment Guidelines
High frequency
RF emissions in
accordance with
CISPR 11
Group 1 The device uses high frequency energy exclu-
sively for its internal functions. For this reason,
its high frequency RF emission is very low and
it is improbable that neighboring electronic
equipment will be disrupted.
High frequency
RF emissions in
accordance with
CISPR 11
Class A
The emissions characteristics of this equipment
make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11
class B is normally required) this equipment
might not offer adequate protection to radio-
frequency communication services. The user
might need to take mitigation measures, such
as relocating or re-orienting the equipment.
Harmonic emis-
sion in accord-
ance with IEC
61000-3-2
Class A
Voltage fluctua-
tions / flickering
in accordance
with IEC
61000-3-3
Fulfilled
The device is used in a professional healthcare / radiological environment.
Environmental conditions are stated in the user manual.
This device was tested for a professional healthcare environment as described
above. Nevertheless the HF-emission and immunity can be influenced by
connected data cables depending on length and the manner of installation.
This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such
an environment.
Resistance to
Jamming Test
IEC 60601-1-2:2014
Test Level
Level of
Agreement
Electromagnet-
ic Environment
Guidelines
Discharge of
static electricity
+ 8 kV contact dis-
charge
+ 8 kV contact
discharge
Floors should
consist of wood,
Crystal-M Imager | Remarks for HF-emission and immunity |
129
3852B EN 20210601 1655
Содержание 5367/300
Страница 1: ...Crystal M Imager 5367 300 User Manual 3852B EN 20210601 1655 ...
Страница 12: ...12 Crystal M Imager Crystal M Imager 3852B EN 20210601 1655 ...
Страница 57: ... The display Crystal M Imager Introduction 57 3852B EN 20210601 1655 ...
Страница 89: ...WARNING Never reuse a jammed film Crystal M Imager Basic Operation operator mode 89 3852B EN 20210601 1655 ...
Страница 128: ...5 Press the OK button on the touchscreen 128 Crystal M Imager Preventive maintenance schedule 3852B EN 20210601 1655 ...
Страница 134: ... Cables transducers and accessories 134 Crystal M Imager Remarks for HF emission and immunity 3852B EN 20210601 1655 ...
Страница 152: ...4 18 2832E EN 20151008 5 Remove plastic bag 152 Crystal M Imager Plug Play Installation Manual 3852B EN 20210601 1655 ...
Страница 161: ...Crystal M Imager Plug Play Installation Manual 161 3852B EN 20210601 1655 ...