Daily Density
Control Chart
Imager
Type: __________ Serial #: _____________ Film Type:____________
Emul #:___________
Densitometer
Internal: ____________ (default selection)
Da te :
Initia ls :
Low Density
Mid Density
Upper Control limit =
+0.05
Calculated Reference
Low Density Level
Lower Control Limit =
-0.05
Upper Control limit =
+0.15
Calculated Reference
Mid Density Level
Lower Control Limit =
-0.15
General radiography applications
Input Tray: ________
Quality Control for
Chart 2A
Figure 12: Chart 2A, Daily Density Control Chart
140
| Crystal-M Imager | Quality Control Charts
3852B EN 20210601 1655