Quality Control
In order to establish and maintain consistent image quality, a regular
evaluation of image quality is advised.
The printer contains an automatic QC feature that has been designed to
comply with the grayscale reproduction constancy test, according to the
international standard IEC 1223-2-4.
Local Regulations may require other procedures.
The QC procedure consists of two main steps:
• Before initial use, establishing a number of reference values that will be
used for further follow-up and verifying initial image quality.
• After establishing these values, performing regular daily, weekly and
annual quality tests.
Note
:
Repeat the two main steps of the Hard Copy quality
procedures for each input tray loaded with general radiography
films.
The results of these tests are recorded on Quality Control Charts.
The QC image has several additional fields where the QC data can be filled in.
This image should be filed as part of the QC procedure.
Related Links
Establishing the reference values and verifying image quality
on page 94
Performing quality control (QC) tests
on page 102
on page 95
on page 124
Topics:
•
Establishing the reference values and verifying image quality
•
•
Establishing the daily operating reference density levels
•
Establishing the image geometry reference values
•
Verifying Acceptable Spatial Resolution, Artifact Levels and Low
Contrast Visibility
•
Performing quality control (QC) tests
DRY 1.4, DRY 2.4 | Advanced Operation (key-operator mode) |
93
3631B EN 20210601 1655
Содержание 5366/500
Страница 1: ...DRY 1 4 DRY 2 4 5366 510 5366 500 User Manual 3631B EN 20210601 1655 ...
Страница 12: ...12 DRY 1 4 DRY 2 4 DRY 1 4 DRY 2 4 3631B EN 20210601 1655 ...
Страница 57: ... The display DRY 1 4 DRY 2 4 Introduction 57 3631B EN 20210601 1655 ...
Страница 89: ...WARNING Never reuse a jammed film DRY 1 4 DRY 2 4 Basic Operation operator mode 89 3631B EN 20210601 1655 ...
Страница 114: ...5 Press the OK button on the touchscreen 114 DRY 1 4 DRY 2 4 Preventive maintenance schedule 3631B EN 20210601 1655 ...
Страница 120: ... Cables transducers and accessories 120 DRY 1 4 DRY 2 4 Remarks for HF emission and immunity 3631B EN 20210601 1655 ...
Страница 132: ...4 18 2832E EN 20151008 5 Remove plastic bag 132 DRY 1 4 DRY 2 4 Plug Play Installation Manual 3631B EN 20210601 1655 ...
Страница 141: ...DRY 1 4 DRY 2 4 Plug Play Installation Manual 141 3631B EN 20210601 1655 ...