4.
Compare the measured A and B dimensions with the reference dimension
values, A
ref
and B
ref
on Chart 4 (‘Geometric Consistency Control Chart’).
The differences between measured dimensions of A and B and the
reference values A
ref
and B
ref
should be less than or equal to 1.0%.
5.
Check for image distortion.
6.
Calculate the aspect ratio by dividing A by B.
The result must be 1 +/- 0.01
WARNING:
If the image size or distortion values are outside of limits,
contact your local service organization to resolve the problem.
Related Links
Establishing the image geometry reference values
on page 100
DRY 1.4, DRY 2.4 | Advanced Operation (key-operator mode) |
105
3631B EN 20210601 1655
Содержание 5366/500
Страница 1: ...DRY 1 4 DRY 2 4 5366 510 5366 500 User Manual 3631B EN 20210601 1655 ...
Страница 12: ...12 DRY 1 4 DRY 2 4 DRY 1 4 DRY 2 4 3631B EN 20210601 1655 ...
Страница 57: ... The display DRY 1 4 DRY 2 4 Introduction 57 3631B EN 20210601 1655 ...
Страница 89: ...WARNING Never reuse a jammed film DRY 1 4 DRY 2 4 Basic Operation operator mode 89 3631B EN 20210601 1655 ...
Страница 114: ...5 Press the OK button on the touchscreen 114 DRY 1 4 DRY 2 4 Preventive maintenance schedule 3631B EN 20210601 1655 ...
Страница 120: ... Cables transducers and accessories 120 DRY 1 4 DRY 2 4 Remarks for HF emission and immunity 3631B EN 20210601 1655 ...
Страница 132: ...4 18 2832E EN 20151008 5 Remove plastic bag 132 DRY 1 4 DRY 2 4 Plug Play Installation Manual 3631B EN 20210601 1655 ...
Страница 141: ...DRY 1 4 DRY 2 4 Plug Play Installation Manual 141 3631B EN 20210601 1655 ...