52
2386G EN 20120110
Operating CR 30-X/CR 30-Xm
CR 30-X/CR 30-X
M
Behavior in Case of Power Failure
In case of a power failure, the system is still connected to the UPS. Two
situations are possible:
•
Power failure after cassette insertion and before identification with the NX
workstation. The digitizer pushes the image plate back in the cassette
without scanning and releases the cassette. After the power supply returns,
the cassette must be inserted into the digitizer and identified again to read
out the image.
•
Power failure after identification with the NX workstation. The image plate
is scanned and erased as usual. The scan cycle finishes when the cassette is
released. If the power supply is still not available, the digitizer will refuse
scanning other cassettes.
Note: The description below is only applicable if an uninterruptible power
supply (UPS) is put into the CR 30-X/CR 30-Xm System configuration. For
more information, refer to the CR 30-X/CR 30-Xm System User Manual,
document 2385.