48
2386G EN 20120110
Operating CR 30-X/CR 30-Xm
CR 30-X/CR 30-X
M
Error during
writing on RF-tag
after scanning
process
“Error during writing on
cassette chip after
successful scan. Erase the
image plate again.”
Red, flashing.
After
unlocking:
green,
constant
Confirm the message
with the OK button on
PC, remove the
cassette, push the
erase button on the
digitizer and re-enter
the cassette for a
manual erase cycle.
Digitizer errors
Erasure lamp
failed during
scanning cycle
“Failure at erasure module.
Please restart digitizer. If
error remains, please
contact service. Image plate
must be erased after repair
again.”
Red, constant
Restart the digitizer or
call service.
Image plate
detected in the
digitizer during
self-test
“Image plate during self-
test in the digitizer
detected. The image plate
was moved back into the
cassette. Please check
whether the image is
present at the PC and re-
scan the cassette, if
necessary. If the image is
present, the cassette was
not erased possibly. Erase
the cassette again in any
case.”
Red, flashing.
After
unlocking:
green,
constant
After power failure a
remaining image
plate, in the digitizer,
will be detected and
the forementioned
error message will be
displayed.
In order to release the
cassette, confirm the
message.
Digitizer cannot
detect the
cassette chip on
the image plate
tray e.g. tray is
missing or was
not inserted
correctly.
“Cassette chip not readable
or cassette tray is missing.
Remove cassette.”
Red, flashing
After
unlocking:
green,
constant
Open the cassette and
check if the tray is
inserted correctly or
use another cassette
and call service.
Condition
Control PC message
Status
indicator
Action
Содержание 5175/200
Страница 1: ...CR 30 X type 5175 200 5175 205 CR 30 Xm type 5179 100 User Manual m 2386G EN 20120110 ...
Страница 53: ...Equipment Information Sheet This appendix covers the following topic T Specifications Appendix A ...
Страница 58: ...58 2386G EN 20120110 Equipment Information Sheet CR 30 X CR 30 XM ...
Страница 59: ...Remarks for HF emission and immunity Appendix B ...
Страница 65: ...65 2386G EN 20120110 CR 30 X CR 30 XM ...
Страница 66: ...0413 Printed in Belgium Published by Agfa HealthCare N V B 2640 Mortsel Belgium 2386G EN 20120110 ...