Mounting
EN
Oxystem 250
5.4
Electrical connection
Verify that the signal cable is not routed immediately next to or together
with cables carrying current.
Verify that the signal cable is sufficiently protected from damage (for
example, by a metal pipe).
DANGER
ELECTRIC SHOCK
•
Verify that the degree of protection against electric shock (protection class,
double insulation) is not reduced by the type of electrical installation.
•
Verify that the product is connected by means of a permanently installed
cable connection.
Failure to follow these instructions will result in death or serious injury.
DANGER
ELECTRIC SHOCK CAUSED BY LIVE PARTS
•
Disconnect the mains voltage supply before performing the work and ensure
that it cannot be switched on.
•
Verify that no hazards can be caused by electrically conductive objects or
media.
Failure to follow these instructions will result in death or serious injury.
NOTICE
UNAVAILABLE MONITORING FUNCTION
•
Do not install mains plugs or switches in the supply line to the product.
•
Only power on/power off the product via the on-site mains fuse.
Failure to follow these instructions can result in equipment damage.