Reference installation, Section B2.1.
1) Steel Wall Plate dimensions are 20” wide x 15 ¾” high x 1 ¼” thick. Wall Mounting bolt
hole pattern is on standard 16” centers horizontally and 7” centers vertically. The hollow,
steel box design and construction allows for internal (hidden) wiring and interconnection of
components.
2) Before installation, review site specifications, table location in the room layout, 120/230 Volt
power supply location and wall mounting location. Respecting local X-ray safety codes,
determine the most suitable location for the operator of the X-Mind Timer Hand Control,
which includes a 12 foot extension coil cord.
3) Determine the best location of the Wall Mounting Plate. Use the Template
(P/N 897-000162) information to determine the reach of the X-ray tube support system
towards the operatory table and patient so sufficient coverage will be achieved.
4) Prepare or plan the most desirable wiring method for the X-Mind Timer in the operatory site
considering the location of an available 120/230 Volt power supply.
5) Mark the six Wall Mounting Plate bolt hole locations on the wall, in alignment with the
internal wall studs or other sufficient support means.
DO NOT MOUNT WALL PLATE OR X-RAY UNIT AT THIS TIME.
6) Determine if external surface wiring or internal to the wall wiring will be used to provide
120/230 Volts to the X-Mind Timer control.
7) Route the 120/230 Volt power supply to a point central to the Wall Plate location using
surface wiring or internal wall wiring. Supply an electrical termination box as per local
codes.
8) If wiring is internal to the wall, allow a 24” pigtail of the 120/230 Volt supply line and ground
to exit from the wall.
9) Using six appropriate 5/16 dia. lag bolts or toggle bolts (DO NOT USE PLASTIC WALL
SHIELDS) fasten the Wall Plate to the wall after routing the 24 inch 120/230 Volt power
supply pig tail to the selected X-Mind Timer wire access hole near the bottom of the plate.
10) Remove the plastic X-Mind cover by loosening the two screws placed on the lower part and
lifting the cover from the bottom to the top to release the upper clamps.
11) Mount the X-Mind Timer Assembly on center of the Wall Plate using three 5/16”-18 x 1”
Allen Cap Screws.
12) Route the power supply wires and to the terminal strip on the bottom of the timer. See
Section B2.2 for detailed electrical connection information.
13) Mount Extension Arm, Scissors Arm and X-ray Tube as per section B2.1
14) Insert six 1”dia. white plastic hole plugs to finish appearance.