23
Anode
Cathode
Transformer insulation
Mineral Oil bath
Exposure time interval
1 s exposure - 30 s pause
Min. focus to skin distance
7.9 in (20 cm)
Diameter of X-ray beam
< 2.36 in (6 cm)
(when it reaches the skin at end of cone)
Cooling
Convection
Radiation leakage at 1 m
< 0.25 mGy/h
Radiation leakage characteristics
70 kV - 8 mA, 1 s @ 132V per cycle,
1 exposure every 30 seconds
Focal spot
0.8 mm (IEC 336)
Anode material
TUNGSTEN (W)
Datum line and anode inclination angle (Heel Angle)
Operating temperature range
+50°F - +104°F (+10°C - +40°C)
Operating relative humidity range
30% - 75%
Temperature range for transport and storage
-4°F - +168°F (-20°C - +70°C)
Max. relative humidity for transport and storage
<95% non-condensing
Min. atmospheric pressure for storage and transport
630hPa
A8. Fuses
ID.
RATING
PARTS PROTECTED
LOCATION
120V
230V
F1
T 10A
T 6.3A
120/230 V Tubehead Protection
Control unit
F2
T 100mA
T 100mA
10 V Secondary winding
Control unit
of transformer
F3
T 1A
T 1A
10 V Secondary winding
Control unit
of transformer
F4
T 10A
T6.3A
120/230 V X-ray, X-ray equip.
Control unit
power supply circuit
NOTE: The “T” rating for fuses stands for time-delay (similar to SLOW-BLOW)