A7. Technical Specifications
The X-ray tubehead has been certified by the manufacturer to be in compliance with FDA
regulations as per 21 CFR SUBCHAPTER J as amended. Periodic compliance testing of
this equipment may be required by local governmental agencies. There are no regulatory
performance standards for veterinary X-ray equipment.
System Classification:
Electromedical Equipment, Class 2 (FDA)
Model Type
Image-Vet 70
ACP
Installation plan
See AFP drawing #897-000162
Supply Ratings:
Line voltage
120/230 VAC +/- 10%
Line frequency
60/50Hz
Maximum absorbed current
7.8 A rms impulsive @ 132 VAC
Power consumption
1.1 kVA impulsive @ 132 VAC
Line voltage regulation
<3%
Circuit type
Single phase self-rectifying with grid control
X-ray System Ratings:
Programmed times
18 selectable times in radiographic mode
Film / Sensor sensitivity
9 levels
Compensated load factors
The exposure time is compensated
on the basis of the input voltage applied
Min. exposure time
0.02 s.
Max. exposure time
1.28 s.
Indicated time resolution
0.02 s.
Preheating time
100 ms
X-ray tube manufacturer and type
CEI - OCX 70-G (with grid)
Toshiba KL16-0.8-70G (with grid)
Tolerances:
Nominal high voltage value
70 kVp + 15%
Anode current
8 mA + 15%
Accuracy of indicated time
+ 48%
Power rating
430 W
Total filtration
2 mm Al eq. @ 70 kV
HVL (Half Value Layer)
>1.5 mm Al eq.
Inherent filtration
(CEI) - 0.5 mm Al eq. @ 70 kV
(KL16) - 0.4 mm Al eq. @ 70 kV