Originale Instructions- Rev.1.3
User Maunal– Loreena
42/ 68
Low threshold:
15.
Leave only a weight of 15 gr in a nacelle.
16.
Start the cycle. 2000 rpm / 1 minute /
9 Acceleration / Braking 9
17.
The centrifuge should achieve the desired speed and then
slow down and stop without the following error message
appearing: The 15 gr imbalance must be tolerated.
Repeat the Low threshold check 3 times: Each cycle should
take place normally.
5.8.
Error Messages, Troubleshooting, Option-Diagnostics
Visual
Instruction
In case of a problem, the display shows an error message and an error
number.
Error / Message
Cause
Solution
01 LOCK FAIL
The lid is open when the centrifuge starts.
Close the lid.
Wait for complete closing of the lid before
starting a new cycle.
The lid sensor is malfunctioning.
Contact a Certified Technician.
02 IMBALANCE
The rotor buckets are not loaded
symmetrically.
Wait until complete stop.
Open the lid.
Load the buckets as shown in manual.
Start a new cycle.
The centrifuge is installed on an uneven
bench.
Install the centrifuge on a flat bench/table, as
described in manual.
The imbalance detector is not well calibrated.
Contact a Certified Technician.
03
BOWL
OVERTEMPERATURE
The temperature in the bowl is higher than
43C; the room temperature is too high.
Wait until complete stop.
Turn the AC or Ventilation system on to
obtain an ideal ambient temperature as
described in manual.
The temperature set in the program is not
adapted.
The settings for speed / temperature are not
compatible with a normal utilization.
The temperature of the samples is too high.
Make sure the samples can be cooled to
down to a temperature between 4C and 37C
during 1 hour
The refrigeration doesn’t work.
Contact a Certified Technician.
The temperature probe is malfunctioning.
Contact a Certified Technician.
The difference between the air temperature
and the over temperature alarm set point is
greater than the set value.
Check the value of the temperature alarm
threshold.
Change program settings: speed and/or
temperature.
04
MOTOR
OVERTEMPERATURE
The motor temperature is too high.
A « no-brakes » stop is occurring.
Wait 30 minutes before opening the lid.
Turn the AC or Ventilation system on to
obtain an ideal ambient temperature as
described in manual.
The centrifuge is being used too intensively
(ventilated model).
Let the centrifuge cool down between cycles.
The motor has been damaged
Contact a Certified Technician.
The temperature sensor on the motor has a
bad connection.
Contact a Certified Technician.
Содержание AFI-C300R
Страница 1: ...User Manual Revision 1 3 English Version...
Страница 63: ...Originale Instructions Rev 1 3 User Maunal Loreena 61 68 Angular rotor NU RA24 2...
Страница 69: ...Originale Instructions Rev 1 3 User Maunal Loreena 67 68 10 EC Conformity Declaration...
Страница 70: ...Originale Instructions Rev 1 3 User Maunal Loreena 68 68 Blank page...
Страница 71: ......