Originale Instructions- Rev.1.3
User Maunal– Loreena
13/ 68
2.
Installation Instructions
2.1.
Shipment
AFI Centrifuge takes every reasonable precaution to assure that your LOREENA Centrifuge arrives without damage. Motor
carriers are carefully selected and shipping cartons have been specially designed to insure your purchase. However,
damage can occur in any shipment and the following outlines the steps you should take on receipt of a LOREENA Centrifuge
to be sure that if damage has occurred, the proper claims and actions are taken immediately.
2.1.1.
Damaged Shipments
Terms are factory, unless stated otherwise. Therefore, it is important to check each shipment before acceptance.
If there is visible damage, the material can be accepted after the driver makes a notation on the consignee's copy of the
freight bill. Then an inspection must be made to verify the claim against the carrier. This inspection is the basis of your
filing the claim against the carrier.
If concealed damage is found it is absolutely necessary to NOTIFY THE FREIGHT AGENT AT ONCE and request an inspection.
Without this inspection, the transportation company may not accept a claim for loss or damage. If the carrier will not
perform the inspection, an affidavit must be prepared stating that he was contacted on a certain date and that he failed to
comply with the request. This along with other papers in the customer's possession will support the claim.
2.1.2.
Handling and transport
The package containing the centrifuge needs to be handled by suitable means (pallet truck, etc.).
Please ensure that the people in charge of handling are qualified to handle the lifting equipment.
Maintain the following conditions during transport:
Temperature: -20 to 50 ° C
Relative humidity: <90%, under the dew point
2.2.
Site preparation
Prepare a stable, clean and levelled support (e.g. laboratory bench). This support must be able to support the weight of the
centrifuge and its use, and be firm enough so as not to
generate or maintain vibration. Such vibrations would
degrade the results of the centrifugation, and may result in an
unbalanced error.
A clearance of 12 inch (300mm) around the centrifuge should
be arranged according to IEC 61010-2-020: No potentially
hazardous material, nobody or no object shall be deposited
within the free space.
Open space around the centrifuge
It is recommended that no less than two people are present
to manually lift the centrifuge onto the laboratory bench.
CAUTION
CAUTION
12”
12”
Access area
12”
Содержание AFI-C300R
Страница 1: ...User Manual Revision 1 3 English Version...
Страница 63: ...Originale Instructions Rev 1 3 User Maunal Loreena 61 68 Angular rotor NU RA24 2...
Страница 69: ...Originale Instructions Rev 1 3 User Maunal Loreena 67 68 10 EC Conformity Declaration...
Страница 70: ...Originale Instructions Rev 1 3 User Maunal Loreena 68 68 Blank page...
Страница 71: ......