Safety Instructions
The operator must observe the following precautions when using the centrifuge:
Ensure stability and perfect levelling of the centrifuge.
Check the correct mounting of the rotor and its accessories before starting a centrifugation cycle.
Check tube resistance at maximum applied centrifugal force: chemical resistance to centrifuged products and
mechanical strength at the centrifugal force applied thereto.
Check the condition of the tubes and remove damaged tubes.
Use only the rotors and accessories authorized by the manufacturer AFI Centrifuge.
Swing-out rotors: install four (or two) buckets; never perform any cycle with missing buckets.
Maintain and control accessories.
Balance the load of the rotor around the axis of rotation.
Respect the maximum density of 1.2 g / ml, in particular in case of a cycle at full speed.
Limit volume in case of excess density.
Install the centrifuge in a ventilated area, on a horizontal rigid support to absorb the vibrations generated by the
centrifuge.
During operation, the centrifuge must not be moved or be subject to impact.
Do not attempt to open the lid while the rotor is spinning.
Do not attempt to neutralize the motorized lid lock
Do not lean over the centrifuge during the spin cycle.
Do not stay within a space of 12 inch around the centrifuge longer than necessary.
Do not leave potentially dangerous materials inside the free space.
Use aerosol-barrier accessories when centrifuging bio hazardous material.
Condensation may form inside the centrifuge when it is moved from a cold environment to a warm environment.
Allow the centrifuge to warm up for two hours before use.
1.6.
Information on the refrigerant fluid used
The "refrigerant" type apparatus contains fluorinated greenhouse gases under the Kyoto Protocol. They should therefore
not be released into the atmosphere. Repairs or disassembly of this appliance should only be carried out by qualified
personnel.
The data corresponding to the type of refrigerant, the Global Warming Potential (GWP) value, and the load are detailed in
§ 1.4.1.Models & Specifications.
Содержание AFI-C300R
Страница 1: ...User Manual Revision 1 3 English Version...
Страница 63: ...Originale Instructions Rev 1 3 User Maunal Loreena 61 68 Angular rotor NU RA24 2...
Страница 69: ...Originale Instructions Rev 1 3 User Maunal Loreena 67 68 10 EC Conformity Declaration...
Страница 70: ...Originale Instructions Rev 1 3 User Maunal Loreena 68 68 Blank page...
Страница 71: ......