47113964.2.023РЭ
22
-
Prepare instruments for operation according to their operation manuals.
-
Power up the digital computer. When a load of an operating system is completed, install the
application software which is included in a complete set of delivery of the transducer.
-
Apply the auxiliary supply to the transducer.
-
Start the program "SetComplex 3.1 (EN)", load a configuration file attached to CD:
- «AET_***_МВ.dat» for AETххх-00 transducer;
- «AET_***_МВ_RTC.dat» for AETххх-10 transducer;
- «AET_***_IEC.dat» for AETххх-01 transducer;
- «AET_***_IEC_RTC.dat» for AETххх-11 transducer.
-
Configurations are to be install:
Area
«RS-485-1»
«RS-485-2»
Baudrate, bps
9600
Stop bits quantity
2
Parity
NO
Device address
1
communication protocol
MODBUS ASCII
Area«Measured parameters»
Registers address
Default set
Register size, byte
2
Resolution factor for every measurand
k1
k2
k3
5000
5000
50000
Group ID
1 (default set)
List «Port»
Any available ones
List «Memory cell size, byte»
2
Switch (three-wire system/ four-
wire system)
Depending on test type
-
Press the «Write» to confirm the installation of the configurations
-
The absence of error messages denotes operability of the transducer.
-
Save the configuration of the transducer on the file to the CD.
-
Close the «SetComplex 3.1 (EN)» software.
3.4.3.6 The following operations should be executed testing of intrinsic error of the four-wire sys-
tem:
-
Complete the operations in accordance with 3.4.3.5 under four-wire system.
-
Start the «ComplexМet 3 EN ». Open the work file at the «Configuration file» menu.
-
Configure appropriate parameters in the “RS-485” Menu and «ModBus» Menu.
-
Configure appropriate nominal voltage and current values in the «Nom. val» menu in accor-
dance with transducer’s model.
-
Press «Start» button in the «ComplexMet 3 EN» window.
-
Apply the nominal input signal to the transducer in accordance with 1.2.2.
-
Let the transducer standby for 10 minutes before taking the first reading to thermally stabilize.
-
Set input signal parameters which are specified in tables by turns:
- 14, 17 for AET100 transducer;
- 15, 17 for AET200 transducer;
- 16, 17 for AET 300 or AET 400 transducers.
-
Read the output code at the «ComplexMet 3 EN» and calculate measurand value by inverse
function of transformation. In case of alternation of the adjacent codes the most deviating value
should be read.