MT-30
Installation, Operation & Maintenance
06225-03H.doc
17 of 47
AeroVironment, Inc
Upper voltage limit: Vdc
Lower voltage limit:
Vdc
Upper power limit:
Watts
Lower power limit:
Watts
Lower current limit: Adc
Upper current limit:
Adc
Will this test be controlled locally or remotely?
Locally
Remotely
If remote operation, what test script will be used?
What control mode is required?
Voltage Mode
Current Mode
Power Mode
Combination of above modes
4.1.1 Choosing a Load Configuration
Some loads and test plans may require a particular configuration of the MT-30
load connections. For example, if the test requires current value over 330
Amperes in either direction,
Parallel
configuration must be used.
The Table and Figure 4.1 show the operating limits for the two configurations,
Independent and Parallel. It may be helpful to plot on the graph of Figure 4.1
points that represent the expected load conditions at different times during the
test and then select a configuration that includes all of these points within the
operating limits. Each of the configurations is described in more detail below.
Configuration
Voltage
Current
Power
Independent A
5 to 120 V
±330 ADC
±30 kW
Independent B
5 to 120 V
±170 ADC
±20 kW
Parallel
5 to 120 V
±500 ADC
±30 kW
"
Caution
" Due to the control strategy required to maintain a broad range of accurate