7. Troubleshooting
Series 300 / 305 / 500 / 505
Fault Description
Possible cause
Remedy
No power
Lead connection broken
Power supply failure
Battery flat
S300,305,500 & 505
damaged
Sensor head damaged
Reconnect power lead
Replace 12V DC power supply
Recharge battery
Replace unit
Replace sensor head
External alarm and
control doesn't work
correctly
Incorrect RJ12 pins used
No diode across external
relay coil
Diode polarity incorrect
Check pins and wire correctly
Fit diode across external relay
coil
Fit diode correctly
Sensor failure when
the sensor is new
Insufficient warm up
Air contaminated
Sensor damaged
Run the sensor on full power
for 24-48 hours.
If an O3 source is available,
exposure to 100 ppb for 5
minutes will speed up the
decontamination process
(approximately 30 minutes)
Move the sensor to cleaner
environment and check reading
Replace sensor
Reading high under
zero gas conditions
Background gas level
higher than normal
Interferent gas present
Sensor zero drift
Sensor damaged
Move sensor to clean air and
check reading is zero or close to
zero
Move sensor to clean air and
check reading is zero or close to
zero
ZERO CAL sensor in a clean,
stable background
Replace sensor