AeroClave
RDS 3110 Operator Manual
Released July 1, 2020
Page 6
Use or disclosure of data contained on this sheet is subject to the restriction on the cover page of this User’s Manual.
3.0
PRODUCT DESCRIPTION
3.1
Technical Specifications
Table 3: Technical Specifications for RDS 3110
4.0
SET
-
UP/OPERATION
4.1
Decontamination Set
-
up
4.1.1
Survey the area being decontaminated. Be sure to note the following:
a)
Dimensions: Length, width, and height.
b)
Location of supply and return vents and any other air entry or egress paths, door
jambs, windows, bathroom vents, hoods etc.
c)
Locations of equipment and furniture.
4.1.2
Determine personnel exposure controls. Controls must be in place to prevent unintentional
personnel exposures. Controls can be any or a combination of:
a)
Administrative, where non
-
decontamination personnel are removed from the
area. Shutting down the HVAC system to the area (if possible) is the preferred
isolation approach.
b)
Engineering, where mechanical isolation strategies are employed. Isolation may
be achieved by sealing with tape and plastic.
Model: RDS 3110
SPECIFICATIONS
Voltage
110VAC
Dimensions
19.7” L x 12” W x 18” D
Total Weight
48 lbs.
# of Static Heads
2
# of SmartPorts
1
Bottle/Reservoir
1.2 Gal, 4 Liter