AC Clamp-on Meter Model CM605
CHAPTER 3
SPECIFICATIONS
3.1 Electrical Specifications
Reference Conditions: Accuracy given @ 23°C ± 5°C, 80% RH Max, Conductor centered in jaws, 45-65Hz, No
DC Component.
AC Current ( )
Range
Resolution
Accuracy
Frequency
10A
1mA
2.0% R ± 10cts
50 to 500Hz
100A
10mA
80A: 2.0% R ± 10cts
>80A: 3.5% R ± 10cts
DC Current ( ) - Positive only*
Range
Resolution
Accuracy
10A
1mA
2.5% R ± 10cts
100A
10mA
80A: 2.5% R ± 10cts
>80A: 4.5% R ± 10cts
AC Volts ( )
Range
Resolution
Accuracy
Impedance
Frequency
600V
100mV
1.5% R ± 5cts
10M
Ω
, <50pF
40 to 500Hz
Overload Protection:
660Vrms
DC Volts ( ) - Positive only*
Range
Resolution
Accuracy
Input Impedance
600V
100mV
1.0% R ± 2cts
10M
Ω
Overload Protection:
660Vrms
* For negative measurement, add 2cts to the accuracy