background image

AC Clamp-on Meter Model CM605 

19

 

Limited Warranty

The Model CM 605 is warranted to the owner for a period of one year from the 

date of original purchase against defects in manufacture. This limited warranty 

is given by AEMC

®

, not by the distributor from whom it was purchased. This 

warranty is void if the unit has been tampered with, abused or if the defect is 

related to service not performed by AEMC

®

For full and detailed warranty coverage, go to www.aemc.com. The war-

ranty information is located in our customer service section.
What AEMC

®

 will do:

If  a  malfunction  occurs  within  the  one-year  period,  you  may  return  the 

instrument to us for repair, provided you submit a proof of purchase. AEMC

®

 

will, at its option, repair or replace the faulty material.

Warranty Repairs

What you must do to return an Instrument for Warranty Repair: 

First, request a Customer Service Authorization Number (CSA#) by phone 

or by fax from our Service Department (see address below), then return the 

instrument along with the signed CSA Form. Please write the CSA# on the 

outside of the shipping container. Return the instrument, postage or shipment 

pre-paid to:

 

AEMC

®

 Instruments

 

Service Department 

 

15 Faraday Drive • Dover, NH 03820 USA

 

Tel:  (800) 945-2362  (Ext. 360)

 

 

(603) 749-6434  (Ext. 360)

 

Fax:  (603) 742-2346 or (603) 749-6309

Caution:

 To protect yourself against in-transit loss, we recommend you insure 

your returned material.

NOTE: 

You must obtain a CSA# before returning any instrument.

Содержание CM605

Страница 1: ...AC DC LOW CURRENT CLAMP ON METER CM605 E N G L I S H User Manual...

Страница 2: ...the instrument to our repair and calibration facility for a nominal charge The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer For r...

Страница 3: ...ns 6 3 2 Environmental Specifications 7 3 3 Mechanical Specifications 8 3 4 Safety Specifications 8 4 OPERATION 9 4 1 Preparation Before Use 9 4 2 AC Current Measurement 10 4 3 DC Current Measurement...

Страница 4: ...ed the protection limit values indicated in the specifica tions for each type of measurement Never measure current while the test leads are connected to the input jacks Use extreme caution when measur...

Страница 5: ...mage may result Risk of electric shock The voltage at the parts marked with this symbol may be dangerous 1 2 Receiving Your Shipment Upon receiving your shipment make sure that the contents are consis...

Страница 6: ...s for crowded wiring or areas 45 12mm 2 Safety barrier anti slip guard 3 Lever for jaw opening closing 4 LCD display 5 COM Black and Positive Red input terminal jacks 6 Data hold and Peak hold buttons...

Страница 7: ...ars and the measurement is locked To deactivate press the button again Peak Hold Function This function displays the peak value To activate press this button during a measurement until the symbol appe...

Страница 8: ...0 to 500Hz 100A 10mA 80A 2 0 R 10cts 80A 3 5 R 10cts DC Current Positive only Range Resolution Accuracy 10A 1mA 2 5 R 10cts 100A 10mA 80A 2 5 R 10cts 80A 4 5 R 10cts AC Volts Range Resolution Accuracy...

Страница 9: ...max Polarity When a negative signal is applied the symbol appears Overload If the range is exceeded the symbol is displayed Low Battery Indicator is displayed when the voltage supplied by the batterie...

Страница 10: ...Specifications Maximum Cable Diameter 0 45 12mm Maximum Jaw Opening 0 60 15mm Dimensions 7 44 x 2 80 x 1 46 189 x 71 x 37mm Weight 6 5 oz 180g 3 4 Safety Specifications IEC EN 61010 1 600V CAT II 300V...

Страница 11: ...with pollution degree level 2 for voltages no greater than 300V in relation to the earth Definition of the installation categories see IEC EN 61010 1 publication CAT II CAT II circuits are household...

Страница 12: ...p the jaws around the conductor to be measured Take the current reading after it stabilizes WARNING Immediately remove the test probes from the circuit under test if overload OL is displayed ZERO OFF...

Страница 13: ...ay may be zeroed Press the ZERO button to zero the read ing Press the lever to open the jaws Clamp the jaws around the conductor to be measured If reading is unstable and is hard to read push the HOLD...

Страница 14: ...the red lead to the red input jack and the black lead to the black COM input jack Bring the test probe tips into contact with the test points Take the voltage reading after it stabilizes WARNING Imme...

Страница 15: ...the switch to the range Insert the red lead to the red input jack and the black lead to the black COM input jack Take the voltage reading after it stabilizes WARNING Immediately remove the test probes...

Страница 16: ...t that all capacitors in the measured circuit be fully discharged Turn the rotary switch to the range Insert the red lead to the red input jack and the black lead to the black COM input jack Bring the...

Страница 17: ...and the black lead to the black COM input jack Bring the test probe tips into contact with the sample under test The buzzer sounds when the circuit to be checked is DC or has a resis tance of less th...

Страница 18: ...put jack Connect the ends of both cables to an oscilloscope or another multimeter Clamp the jaws around the conductor to be measured Read the analog signal on the oscilloscope or multimeter If the sig...

Страница 19: ...and or damage to the instrument do not get water or other foreign agents into the probe 5 2 Battery Replacement The batteries will need to be replaced when the symbol appears on the display The meter...

Страница 20: ...bration traceable to N I S T Includes calibration certificate plus recorded calibration data Ship To AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Tel 800 945 2362 Ext 360 603 749 6434 Ext 360...

Страница 21: ...instrument to us for repair provided you submit a proof of purchase AEMC will at its option repair or replace the faulty material Warranty Repairs What you must do to return an Instrument for Warranty...

Страница 22: ...20 AC Clamp on Meter Model CM605 Notes...

Страница 23: ......

Страница 24: ...11 09 99 MAN 100346 v1 Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 603 749 6434 Fax 603 742 2346 www aemc com...

Отзывы: