Infrared K-Thermometer Model CA876

CHAPTER 4

OPERATION

4.1  Infrared Measurement Considerations

MEASUREMENT THEORY

Every object emits infrared (IR) energy proportional to its temperature. By 

measuring the amount of this radiant energy, it is possible to determine 

the temperature of the emitting object. Infrared radiation is invisible light 

(electromagnetic radiation), which easily travels through air and is easily 

absorbed by solid matter. An IR thermometer, which operates by detecting 

infrared radiation, can accurately measure an object surface temperature 

without touching it and independently of the air temperature or the mea-

surement distance.
Infrared radiation, which is emitted from the object, is focused into an infra-

red radiation sensor through an optical system. This system includes an 

optical lens, which is transparent to infrared radiation, and a 5.3μm cut 

off filter. The output signal from the infrared radiation sensor is input to an 

electronic circuit, along with the output signal from a standard temperature 

sensor, to calculate the temperature and display it on the meter display.

EMISSIVITY

All objects emit invisible infrared energy. The amount of IR energy emitted 

is proportional to the object’s temperature and its natural ability to emit IR 

energy. This ability, called emissivity, is based upon the object material 

type and its surface finish. Emissivity values range from 0.10 for a very 

reflective object to 1.00 for a black body. Factory set emissivity value of 

0.95 will cover 90-95% of typical applications.
If frost or other material/substance covers the measured surface, clean it 

to expose the surface. If the surface to be measured is highly reflective, 

apply dull masking tape or matte black paint over the surface. If the ther-

mometer seems to be giving incorrect readings, check the front sensor. 

There may be condensation or debris obstructing the sensor. Only clean 

per instructions.

Содержание CA872

Страница 1: ...INFRARED K THERMOMETER WITH LASER CA876 E N G L I S H User Manual Auto OFF...

Страница 2: ...f the instrument Please record this number and purchase date for your records INFRARED THERMOMETER MODEL CA872 CATALOG 2121 31 SERIAL _ _______________________________________________ PURCHASE DATE __...

Страница 3: ...7 3 1 Environmental Specifications 7 3 1 1 Infrared 7 3 1 2 K Type 8 3 2 General Specifications 9 3 3 Safety Specifications 9 4 OPERATION 10 4 1 Infrared Measurement Considerations 10 4 2 Recommendat...

Страница 4: ...nt errors do not expose the thermom eter to strong electrical or electromagnetic fields To prevent sensor damage and erroneous readings do not put the sensor lens into contact with any body or foreign...

Страница 5: ...the laser beam Laser Output 0 5mW 670nm wavelength 1 2 Receiving Your Shipment Upon receiving your shipment make sure that the contents are consis tent with the packing list Notify your distributor o...

Страница 6: ...ntact when in the IR mode non contact temperature mea surement An aiming laser enables the user to pinpoint the target The K type thermocouple input may also be used The thermometer includes a program...

Страница 7: ...unction is programmable ALM Hi ALM Lo Emissivity may be adjusted with the and buttons to match a particu lar target We recommend leaving it at 0 95 for general use ALM Hi The Hi alarm set point is adj...

Страница 8: ...n turns the laser ON or OFF The symbol is displayed in the upper left hand corner when the laser is ON Remove the laser cover before use The laser is activated when the MEAS button is pressed during a...

Страница 9: ...00 C 2 of Reading or 6 F 3 C whichever is greater 64 4 to 82 4 F 18 to 28 C ambient operating temperature Temperature Coefficient Changes in accuracy operating temperature above 82 4 F 28 C or below 6...

Страница 10: ...C or 1 F C Accuracy 328 F to 1999 F 0 1 of Reading 2 F plus T C 199 C to 1370 C 0 1 of Reading 1 C plus T C Accuracy is specified for operating temperatures over the range of 64 to 82 F 18 to 28 C fo...

Страница 11: ...h battery removed Auto Power Off 15 seconds approx Altitude 2000m max Input Protection 24VDC or 24rms maximum input voltage on any combi nation of inputs Battery Standard 9V battery NEDA 1604 6LR61 or...

Страница 12: ...nd a 5 3 m cut off filter The output signal from the infrared radiation sensor is input to an electronic circuit along with the output signal from a standard temperature sensor to calculate the temper...

Страница 13: ...80 to 0 90 Copper oxide 0 78 Mortar 0 89 to 0 91 Iron oxide 0 78 to 0 82 Brick red 0 93 to 0 96 Textiles 0 90 4 2 Recommendations before Operating If the measured surface target diameter is less than...

Страница 14: ...pressed When the button is released the measurement will be held in the display for 15 20 seconds HOLD is displayed in the lower left hand corner of the display 8 The IR thermometer will shut OFF auto...

Страница 15: ...ermometer is OFF hold down the button then press the MEAS button F will be displayed The selected temperature scale will remain until changed by the user 4 6 Continuous Measurement The user may want t...

Страница 16: ...e LCD display when replacement is needed Replace with a standard 9 volt alkaline battery NEDA 1604 6LR61 To replace the battery Turn the meter OFF Remove the rubber holster Remove the screw from the b...

Страница 17: ...ncludes calibration certificate plus recorded calibration data Ship To Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 800 945 2362 Ext 360 603 749 6434 Ext 360 Fax...

Страница 18: ...ay return the instrument to us for repair provided we have your warranty registration information on file or a proof of purchase AEMC Instruments will at its option repair or replace the faulty materi...

Страница 19: ...Notes...

Страница 20: ...06 04 99 MAN 100246 v4 Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 603 749 6434 Fax 603 742 2346 www aemc com...

Отзывы: